Process-Related Faults And Messages - Miele PG 8581 Operating Instructions Manual

Dental washer-disinfector
Table of Contents

Advertisement

Frequently asked questions
Problem
Fault 550
Fault 555
Fault 559
Fault 578

Process-related faults and messages

Problem
Drying during program
deactivated
Wrong code entered
114
Possible cause and solution
The waterproof system has been activated. One of the
water supply hoses might have a leak.
Close the water faucets.
Contact Miele Service.
Too much water has accumulated in the steam condenser.
Restart the machine. Excess water is pumped out
automatically.
There is a problem with the process documentation
interface. The machine has detected a module for and
Ethernet interface, but only a serial interface is activated in
the controls (RS232).
Deactivate the RS232 interface:
Open the menu for configuring the interface
settings
Interface
/
Wait approx. 90 seconds. The Ethernet module XKM
3000 L Med needs this time for initialization. It may be
necessary to reconfigure the interface.
Or
Replace the Ethernet module XKM 3000 L Med with a
XKM RS232 10 Med module to set up a serial interface.
The peak-load cut-out has lasted longer than 3 hours.
Have your electrical system and your energy
management system tested by a suitably qualified
person.
Possible cause and solution
Drying cannot be selected at the start of a program
because drying is not available for the selected program.
Start the program without drying.
or
Have the drying parameters for this program adjusted by
Miele Service.
The PIN code entered is not the same as the code saved.
Enter the PIN code again.
Report the loss of the PIN code to Miele Service.