Repeating if limit
value is exceeded
Calibrating the
conductivity
meter
If the conductivity exceeds the limit value for the water intake or
drainage, the measurement can be repeated.
If the limit value of the water intake is too high, the water is pumped
out and the water intake is repeated. If the conductivity at the end of
the rinse cycle is above the limit value for the drainage water, the
entire final rinse is repeated.
If the measured value is still above the limit following the repetition,
the programme is cancelled with a error message. The error message
is shown in the display and also recorded in the process
documentation.
...
LFMMc max. value
Water intake
Number of repeats
Adjustment is in increments of 1. The possible range is shown in the
bottom line.
Use the arrow buttons (higher) and (lower) to set the number of
repetitions.
Press OK to save the setting.
The conductivity meter must be recalibrated by Miele Service at
regular intervals, e.g. during servicing. Sometimes calibration may
also be necessary outside the service cycle.
Conductivity module
requires calibration
If this message appears, contact the Miele Service Department.
Water drainage
or
Number of repeats
1
(0 - 1)
Programme settings
107