Adding and dispensing chemical agents
Replenishing
liquid cleaning
agent
52
Open the drawer in the side unit.
Remove the liquid cleaning agent container (blue marking) and
place it on the open cabinet door or on a surface which is robust
and easy to clean.
Take the lid off the canister and remove the suction lance. Place the
suction lance on the open wash cabinet door.
Replace the empty container with a full one.
Push the suction lance into the opening of the container and secure
the lid. Observe the colour coding.
Feed the suction lance into the container until it reaches the
bottom.
Wipe up any spilled process chemical thoroughly.
Place the container back in the drawer in the side cabinet.
Close the drawer. Ensure that the dispensing tubes and cables are
not kinked or trapped.
The dispensing system must then be vented (see "Settings /DOS
venting").