Release Programme - Miele PG 8583 CD Operating Instructions Manual

Washer-disinfector for laboratory glassware and laboratory utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Release programme

It is possible to block access to individual programmes. Blocked
programmes are not available for selection, so for example it can be
ensured that only validated programmes are used.
Open the menu as follows:
 button
     
Further settings
Release programme
         
Release programme
All
Selection
-
All
All programmes are released for use.
Selection
-
A selection of programmes are available for use.
Select an option using the  and  arrow buttons and confirm your
selection with OK.
Selection
The
option displays a list of all programmes.
Release programme
Accept
Programmes are selected by multiple choice. A box is shown next to
all programmes in the list. If a programme is released, there is a tick
 in the box. An empty box indicates a blocked programme.
Programmes can be released or blocked using the arrow buttons 
and  and by confirming with OK.
To save the selection select the
and confirm with OK.
Universal
Accept
Further settings
option at the end of the list
83