Miele PG 8583 CD Operating Instructions Manual page 142

Washer-disinfector for laboratory glassware and laboratory utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Problem solving guide
Problem
Test programme: test object
can now be removed
Programme cancelled
Programme continued
Peak load cut-out
All settings reset
All programme settings reset
142
Cause and remedy
This is not a fault.
A test programme is running to check performance. At
certain points in the programme the sequence is
interrupted so that samples can be taken.
Take a sample.
or
Wait. The programme will continue automatically in
approx. 30 seconds.
or
Continue the programme without delay by pressing the
Start/Stop button.
This is not a fault.
A programme which was running was cancelled by the
user.
 The wash cabinet interior can be very hot.
When the door is opened, hot steam and chemical
agents can escape. Protective measures for personal
safety must be observed.
This is not a fault.
The process of cancelling a programme was not
completed.
The programme which was running continued without
interruption.
This is not a fault.
Individual components of the machine are paused while
there is a peak load signal from your energy management
system.
This is not a fault.
A user has restored factory default settings.
Confirm the message with OK.
This is not a fault.
A user has restored the factory setting for the programme.
Confirm the message with OK.