Kikusui BIM1000 Series User Manual page 82

Battery impedance meter
Table of Contents

Advertisement

Specifications | Basic specifications
Resistance meter
Item
Measurement method
Range
Maximum display
*1
value
Resolution
Measured current
Measurement frequency
*3
Accuracy
Temperature coeffi-
cient
*4
Response time
*1.
Displays OVER when the measurement range is exceeded.
*2.
Within error ± 10 %.
*3.
Add ±3 digit when the sampling speed is set to FAST and ±2 digit when the sampling speed is set to MID.
*4.
The time for the product's measurement circuit to stabilize when a probe in an open state is connected to the DUT.
Sampling time
Item
Sampling speed
Judgment function
Item
Judgment method
Resistance
Voltage
82
User's Manual
3 mΩ range
30 mΩ range
300 mΩ range
3 Ω range
3 mΩ range
30 mΩ range
300 mΩ range
3 Ω range
*2
3 mΩ range
30 mΩ range
300 mΩ range
3 Ω range
3 mΩ range
30 mΩ range
300 mΩ range
3 Ω range
Power supply frequency 50 Hz
Power supply frequency 60 Hz
BIM1030
Window comparator method. Judgment made with software.
Setting range
0.0000 Ω to 3.1000 Ω
Resolution
100 µΩ
Setting range
0.000 V to 315.000 V
Resolution
1 mV
Specifications
Four-terminal measurement method
3 mΩ, 30 mΩ, 300 mΩ, 3 Ω, AUTO
3.1000 mΩ
31.000 mΩ
310.00 mΩ
3.1000 Ω
0.1 μΩ
1 μΩ
10 μΩ
100 μΩ
100 mA
100 mA
10 mA
1 mA
1 kHz ±0.2 Hz
±(0.5 % of reading + 5 digit)
±(0.05 % of reading + 1 digit) /°C
±(0.05 % of reading + 0.5 digit) /°C
±(0.05 % of reading + 0.5 digit) /°C
±(0.05 % of reading + 0.5 digit) /°C
Approx. 2 ms
Specifications
FAST: 20 ms, MID: 50 ms, SLOW: 160 ms
FAST: 20 ms, MID: 42 ms, SLOW: 150 ms
BIM1100
0.00 V to 1050.00 V
10 mV
BIM1000 Series

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bim1030Bim1100

Table of Contents