Event status register 0 (ESR0)
Bit
Bit
weight
0
1
1
2
2
4
3
8
4
16
5
32
6
64
7
128
Event status register 1 (ESR1)
Bit
Bit
weight
0
1
1
2
2
4
3
8
4
16
5
32
6
64
7
128
BIM1000 Series
Signal
Description
name
EOM
Set when a measurement is completed.
INDEX
Set while a measurement is being performed.
–
Not used
–
Not used
–
Not used
ERR
Set when a measurement error occurs.
–
Not used
–
Not used
Signal
Description
name
R-Low
Set when the resistance measurement judgment result is LOW.
R-IN
Set when the resistance measurement judgment result is IN.
R-Up
Set when the resistance measurement judgment result is UP.
V-Low
Set when the voltage measurement judgment result is low.
V-IN
Set when the voltage measurement judgment result is IN.
V-Up
Set when the voltage measurement judgment result is UP.
PASS
Set when the resistance and voltage measurement judgment results are both IN.
FAIL
Set when resistance and voltage measurement judgment results include LOW or
UP.