Kikusui BIM1000 Series User Manual page 60

Battery impedance meter
Table of Contents

Advertisement

Remote Control | Message Reference
*CLS
Clears all status byte registers and event status registers.
*ESE
Sets or queries the standard event status enable register (SESER). Set using a total value of the weight
(p. 52)
of the bits you want to set.
Program header
Value/Response
Example
*ESR?
Queries and then clears the standard event status register (SESR). The set bits are returned using a bit
weight
(p.
Program header
Value/Response
Example
*SRE
Sets or queries the service request enable register (SRER). Set using a total value of the weight
the bits you want to set.
Program header
Value/Response
Example
*STB?
Queries the contents of the status byte register (STB) and the MSS message. The set bits are returned
using a bit weight
Program header
Response
Example
60
User's Manual
*ESE
*ESE?
<0 to 255 (NR1)>
ESE 96
Set SESER bit 6 (bit weight 64) and bit 5 (bit weight 32) to 1.
52).
*ESR?
<0 to 255 (NR1)>
ESR? 32
SESR bit 5 (bit weight 32) is 1.
*SRE
*SRE?
<0 to 255 (NR1)>
SRE 17
SESR bit 0 (bit weight 1) and bit 4 (bit weight 16) are 1.
(p.
51).
*STB?
<0 to 255 (NR1)>
32
STB bit 5 (bit weight 32) is 1.
(p. 51)
of
BIM1000 Series

Advertisement