Remote Control | Message Reference
Reading errors
Retrieves the oldest error from the error queue. Errors that have been read are cleared. The error/event
queue can hold up to 16 errors. If the 17th error occurs, the 16th error changes to "-350 Queue overflow,"
and subsequent errors are not stored.
The error queue is cleared when the product is turned off, when the last entry in the queue is read, or
when a *CLS command is received.
Program header
Response
(Errors)
Restoring the factory default settings
Resets the product settings to their factory defaults
Program header
Note
70
User's Manual
:SYSTem:ERRor?
+0, No error
There are no errors.
-100, Command error
There is a command error.
-310, System error
There is a system error.
-350, Queue overflow
The error could not be placed in the queue because the queue
is full.
:SYSTem:RESet
If you do not want to delete the measurement conditions saved to memory and RS232C
settings, use *RST.