Retrieving the most recent measured value
Queries the most recent measured value.
Program header
Response
Unit
Note
 Example of retrieving measured values using :FETCh?
If :INITiate:CONTinuous
send "*TRG" or SIGNAL I/O TRIG-IN signal or press SNGL TRG on the front panel, the product makes a
single measurement. When you send :FETCh?, the measured value is read.
Executing a measurement and retrieving the measured value
When continuous measurement (:INITiate:CONTinuous) is off, a query for the measured value is made
after executing a measurement once.
When the trigger source (:TRIGger:SOURce) is set to EXT, a measurement is not executed by this com-
mand alone. To execute a measurement, use SNGL TRG on the front panel or the SIGNAL I/O TRIG-IN
signal (a measurement cannot be performed using *TRG).
Program header
Response
Unit
Note
 Example of retrieving measured values using :READ?
If :INITiate:CONTinuous
:READ? causes the measured value to be read after a measurement is made.
If :INITiate:CONTinuous
:READ? causes the product to switch to a trigger wait state. If you send a SIGNAL I/O TRIG-IN signal, the
product makes a single measurement, and the measured value is read.
BIM1000 Series
:FETCh?
Any of the following responses will be returned depending on the :DATA:FORMat
setting.
Resistance only: <Resistance measurement (NR3)>
Voltage only: <Voltage measurement (NR3)>
Resistance and voltage: <Resistance measurement (NR3)>,<Voltage measurement
(NR3)>
Ω, V
When the measurement overflows, "+1.00000+E08" is returned. When a measurement
error occurs, "+1.00000+E09" is returned.
(p. 72)
is set to ON and :TRIGger:SOURce
:READ?
Any of the following responses will be returned depending on the :DATA:FORMat
setting.
Resistance only: <Resistance measurement (NR3)>
Voltage only: <Voltage measurement (NR3)>
Resistance and voltage: <Resistance measurement (NR3)>,<Voltage measurement
(NR3)>
Ω, V
When continuous measurement (:INITiate:CONTinuous) is on, an execution error
occurs.
The next command is not executed until the measurement is completed.
To execute this command after changing the measurement conditions, insert an interval
of 100 ms.
When the measurement overflows, "+1.00000+E08" is returned. When a measurement
error occurs, "+1.00000+E09" is returned.
(p. 72)
is set to OFF and :TRIGger:SOURce
(p. 72)
is set to OFF and :TRIGger:SOURce
Remote Control | Message Reference
(p. 72)
is set to EXTernal, when you
(p. 72)
is set to INTernal, sending
(p. 72)
is set to EXTernal, sending
User's Manual
(p. 74)
(p. 74)