Kikusui BIM1000 Series User Manual page 63

Battery impedance meter
Table of Contents

Advertisement

Voltage measurement range
When you send a voltage value you want to measure, the range is set appropriately for measuring that
voltage.
Program header
Value
BIM1030
BIM1100
Unit
Response
Example
Note
Auto measurement range
Program header
Value
Response
Zero adjustment
Executes a zero adjustment and queries the result. Zero adjustment takes time. About 10 seconds are
required to the next process.
Program header
Response
 Clears zero adjustment
Program header
BIM1000 Series
:VOLTage:RANGe
:VOLTage:RANGe?
<0 to 315 (NR2)>
<-1050 to 1050 (NR2)>
V
<NR3>
Send ": VOLT: RANG 500"
The range is set to 600 V, which allows 500 V to be measured.
Send "VOLT: RANG?"
The present range setting "6.00000E+02" (600 V) is returned.
If :AUTorange is set to ON (1) or VOLT, the value changes automatically.
:AUTorange
:AUTorange?
{1|0|ON|OFF|RESistance|VOLTage}
{1|ON}:
Sets the resistance and voltage measurement range to auto
{0|OFF}:
Sets the resistance and voltage measurement range to manual
RESistance:
Sets the resistance measurement range to auto
VOLTage:
Sets the voltage measurement range to auto
{ON|OFF|RESISTANCE|VOLTAGE}
:ADJust?
{0|1}
0: Zero adjustment complete
1: Zero adjustment failure
:ADJust:CLEAr
Remote Control | Message Reference
User's Manual
63

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bim1030Bim1100

Table of Contents