Kikusui BIM1000 Series User Manual page 21

Battery impedance meter
Table of Contents

Advertisement

5
Press Do 0 Adj (F1).
Zero adjustment starts. When zero adjustment is complete, a dialog appears showing the adjust-
ment results of each range.
If zero adjustment is successful, "OK" appears. If the allowable zero adjustment range is
exceeded, "FAIL" appears.
6
Press OK (F5) to close the dialog box.
If zero adjustment has been performed, "0 ADJ" is displayed in green at the top area of the range
display.
This completes the setting.
If "FAIL" is displayed
Zero adjustment has a tolerance of up to 1,000 digits. If this tolerance is exceeded during zero adjustment,
"FAIL" will be displayed in the adjustment result dialog box. Check that the cables are connected correctly,
and perform zero adjustment again.
Using the zero adjustment tool (option)
If you are performing zero adjustment using the optional clip-type four-wire test leads
the optional zero adjustment tool
the zero adjustment tool, and perform zero adjustment.
BIM1000 Series
(p.
90). As shown in the following figure, insert the clips into the holes of
Zero adjustment tool
Measurement | Taking Measurements
(p.
89), you can use
User's Manual