Kikusui BIM1000 Series User Manual page 74

Battery impedance meter
Table of Contents

Advertisement

Remote Control | Message Reference
Setting the response format
Sets the :FETCh? and :READ? response format to resistance, voltage, or both.
Program header
Value
Response
Example
Reading the SIGNAL I/O input status
Reads the input status of SIGNAL I/O STB and TRIG IN.
Program header
Response
74
User's Manual
:DATA:FORMat
:DATA:FORMat?
{RESistance|VOLTage|RESistance,VOLTage}
RESistance:
Resistance only
VOLTage:
Voltage only
RESistance,VOLTage: Resistance and voltage
{RESISTANCE|VOLTAGE|RESISTANCE,VOLTAGE}
:DATA:FORMat RESistance,VOLTage
:IO:IN?
{32|33|1|0}
32:
STB is on.
33:
STB and TRIG IN are on.
1:
TRIG IN is on.
0:
STB and TRIG IN are off.
BIM1000 Series

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bim1030Bim1100

Table of Contents