Voltage judgment function
Program header
Value
Response
Note
Upper resistance judgment limit
Program header
Value/Response
Unit
Note
Lower resistance judgment limit
Program header
Value/Response
Unit
Note
Upper voltage judgment limit
Program header
Value/Response
Unit
Note
Lower voltage judgment limit
Program header
Value/Response
Unit
Note
BIM1000 Series
:SENSe:JUDGment:VOLTage:STATe
:SENSe:JUDGment:VOLTage:STATe?
{BOTH|OFF|UPPer|LOWer}
BOTH:
Turns on the upper and lower voltage limit judgment functions simultaneously
OFF:
Turns off the upper and lower voltage limit judgment functions simultaneously
UPPer:
Turns on only the upper voltage limit judgment
LOWer: Turns on only the lower voltage limit judgment
{BOTH|OFF|UPPER|LOWER}
If :SENSe:JUDGment:STATe is set to ON (1), the value changes to BOTH.
:SENSe:JUDGment:RESistance:UPPer
:SENSe:JUDGment:RESistance:UPPer?
<NR2>
Ω
Set a value greater than the lower judgment limit.
:SENSe:JUDGment:RESistance:LOWer
:SENSe:JUDGment:RESistance:LOWer?
<NR2>
Ω
Set a value less than the upper judgment limit.
:SENSe:JUDGment:VOLTage:UPPer
:SENSe:JUDGment:VOLTage:UPPer?
<NR2>
V
Set a value greater than the lower judgment limit.
:SENSe:JUDGment:VOLTage:LOWer
:SENSe:JUDGment:VOLTage:LOWer?
<NR2>
V
Set a value less than the upper judgment limit.
Remote Control | Message Reference
User's Manual
65