Remote Control | Message Reference
 Queries the status
Queries the condition of the QUEStionable status register.
A query does not clear the contents of the register.
Program header
Value/Response
QUEStionable status enable register
Sets or queries the QUEStionable status enable register. Set using a total value of the weight
bits you want to set.
Program header
Value/Response
Example
Switching the measurement display
Switches the measured value that is shown on the front panel
Program header
Value
Response
Resistance measurement range
When you send a resistance value you want to measure, the range is set appropriately for measuring that
resistance.
Program header
Value
Unit
Response
Example
Note
62
User's Manual
:STATus:QUEStionable:CONDition?
<0 to 3845 (NR1)>
:STATus:QUEStionable:ENABle
:STATus:QUEStionable:ENABle?
<0 to 3845 (NR1)>
:STATus:QUEStionable:ENABle 1536
Set QUEStionable status enable register bit 9 (bit weight 512) and bit 10 (bit weight
1024) to 1.
:FUNCtion
:FUNCtion?
{RV|RESistance|VOLTage}
RV:
Displays the resistance and voltage
RESistance:
Displays the resistance
VOLTage:
Displays the voltage
{RV|RESISTANCE|VOLTAGE}
:RESistance:RANGe
:RESistance:RANGe?
<0 to 3.1 (NR2)>
Ω
<NR3>
Send ": RES: RANG 0.12"
The range is set to 300 mΩ, which allows 120 mΩ to be measured.
Send ": RES: RANG?"
The present range setting "3.00000E-01" (300 mΩ) is returned.
If :AUTorange is set to ON (1) or RES, the value changes automatically.
(p. 54)
of the