Kikusui BIM1000 Series User Manual page 68

Battery impedance meter
Table of Contents

Advertisement

Remote Control | Message Reference
key lock
Program header
Value
Response
SIGNAL I/O lock
When the SIGNAL I/O lock is set to on, control through SIGNAL I/O will be disabled. This is automatically
turned off when the product is started.
Program header
Value
Response
Switches to local mode
During remote control, the product is in remote mode (no panel operations accepted), and a REMOTE
icon is shown in the display. To change to local mode (panel operations accepted), send the following
command. When the product enters local mode, the REMOTE icon disappears.
Program header
Saving and recalling measurement conditions
 Saving measurement conditions
When you send a command with a designated memory number, measurement conditions is saved to the
specified memory number. Up to 100 entries can be saved to memory.
Program header
Value
Note
 Recalling measurement conditions
When you send a command with a designated memory number, measurement conditions is recalled from
the specified memory number.
Program header
Value
Note
68
User's Manual
:SYSTem:KLOCk
:SYSTem:KLOCk?
{1|0|ON|OFF}
{ON|OFF}
:SYSTem:ELOCk
:SYSTem:ELOCk?
{1|0|ON|OFF}
{ON|OFF}
:SYSTem:LOCal
:SYSTem:SAVE
<0 to 99 (NR1)>
Note that there is a limit to the number of times the conditions can be saved (1 million).
:SYSTem:LOAD
<0 to 99 (NR1)>
If you specify an unsaved memory number, an execution error will occur.
BIM1000 Series