Kikusui BIM1000 Series User Manual page 67

Battery impedance meter
Table of Contents

Advertisement

 Collective log transfer
Program header
Response
Note
Example
Measurement output for when triggers are received
When *TRG is received or TRIG-IN is received through SIGNAL I/O, the measurement is output to the
external device. This eliminates the need to send a measurement query command from the external
device. For details on the measurement that is output, see "Setting the response format"
Program header
Value
Response
Key operation sound
Program header
Value
Response
SCPI command error sound
Program header
Value
Response
Frequency
Program header
Value/Response
BIM1000 Series
:LOG:DATA?
<Save number (NR1)>,<Resistance measurement (NR3)>,<Voltage measurement
(NR3)>
All the logs are read consecutively.
An "END" character is sent at the end of the log.
When :LOG:DATA? is sent
1,1.8348E-1,3.5685E+0
2,1.8354E-1,3.5685E+0
3,1.8360E-1,3.5685E+0
END
:SYSTem:DATAout
:SYSTem:DATAout?
{1|0|ON|OFF}
{ON|OFF}
:SYSTem:BEEPer:STATe
:SYSTem:BEEPer:STATe?
{1|0|ON|OFF}
{ON|OFF}
:SYSTem:SCPI:ERRor
:SYSTem:SCPI:ERRor?
{1|0|ON|OFF}
{ON|OFF}
:SYSTem:LFRequency
:SYSTem:LFRequency?
{50|60}
Remote Control | Message Reference
(p.
User's Manual
74).
67

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bim1030Bim1100

Table of Contents