Kikusui BIM1000 Series User Manual page 69

Battery impedance meter
Table of Contents

Advertisement

Measurement conditions for the next startup
Saves the present measurement conditions and recalls them at the next startup.
Program header
Note
EOM signal output setting
Sets or queries the SIGNAL I/O EOM signal output method.
 EOM signal output mode
Program header
Value
Response
 EOM signal pulse width
Sets the pulse width when :SYSTem:EOM:MODE is set to PULSe.
Program header
Value/Response
Unit
Sets the current pulse output
When the current pulse output is set to PULSE, measurement current flows only during measurements. By
offsetting the trigger inputs, this function prevents measurement errors caused by mutual interference of
measurement currents, such as when measurements are made using several of this product instruments
simultaneously. This is valid when :INITiate:CONTinuous
(p. 72)
is set to EXTernal.
Program header
Value
Response
BIM1000 Series
:SYSTem:BACKup
Note that there is a limit to the number of times the conditions can be saved (1 million).
:SYSTem:EOM:MODE
:SYSTem:EOM:MODE?
{HOLD|PULSe}
HOLD:
When :TRIGger:SOURce
measurement is completed until the next measurement starts. When :TRIG-
ger:SOURce is set to EXTernal, EOM is turned on after a measurement is
completed until the next trigger is detected.
PULSe:
Outputs an EOM signal at the pulse width set with :SYSTem:EOM:PULSe.
{HOLD|PULSE}
:SYSTem:EOM:PULSe
:SYSTem:EOM:PULSe?
<0.001 to 0.099 (NR2)>
Seconds
:SYSTem:CURRent
:SYSTem:CURRent?
{CONTinuous|PULSe}
CONTinuous: Runs measurement current continuously.
PULSe:
Runs measurement current only during measurements. If continuous
measurement
when CONTinuous is specified.
{CONTINUOUS|PULSE}
Remote Control | Message Reference
(p. 72)
is set to INTernal, EOM is turned on after a
(p. 72)
is set to OFF or :TRIGger:SOURce
(p. 72)
is set to on, the operation will be the same as
69
User's Manual