The Analyzer Format Menu

Pod threshold field

The Analyzer Format Menu
Pod threshold field
The pod threshold field is used to set a voltage level that the data must reach
before the analyzer recognizes and displays it as a change in logic levels.
Threshold levels apply to single pods, and cover both data and clock channels.
TTL When TTL is selected as the threshold level, the data signals must
reach +1.5 volts.
ECL When ECL is selected as the threshold level, the data signals must
reach −1.3 volts.
User When User is selected as the threshold level, the data signals must
reach a user selectable value. The range of this value is between −6.0
volts to +6.0 volts in 50-mV increments.
8-32