The Analyzer Format Menu

State acquisition modes (state only)

State acquisition modes (state only)
The State Acquisition Mode field identifies the channel width and memory
depth of the selected acquisition mode. There are two configurations of
channel width/memory depth.
Full Channel/4K Memory/100 MHz
Full-channel mode uses both pods in a pod pair for 34 channels of width and
a total memory depth of 4 K per channel. If time or state tags are turned on,
the total memory is evenly split between data acquisition storage and time or
state tag storage. To maintain the full 4 K per channel depth with state or
time tags, leave one pod pair unassigned.
Half Channel/8K Memory/100 MHz
Half-channel mode cuts the channel width to 17 channels. In this mode, the
pod used within the pod pair is selected through the Pod field. In
half-channel mode, the memory depth is increased to 8 K per channel. Time
or state tags are not available in this mode.
8-33