Ti Interface - Renishaw ATOM Series Installation Manual

Linear encoder systems
Table of Contents

Advertisement

System connection
(continued)

Ti interface

Approved ESD precautions must be
followed at all times during readhead
and interface electrical connection.
The readhead is connected to the Ti interface via
a small, rugged interboard connector to allow for
easy feed-through during installation.
Connecting the readhead
1.
Remove the cover plate as shown
(two M2.5 hex head screws).
2.
Taking care not to touch the pins,
plug the connector into the socket
in the interface, ensuring correct
orientation as shown.
3.
Refit the cover plate, ensuring the cable ferrule is located in the recess on the inside, and no wires are trapped under the cover plate.
4.
Proceed with
'Readhead mounting and alignment', page 23
ATOM linear installation guide
and
'System calibration', page
28.
Disconnecting the readhead
1.
Disconnect power.
2.
Remove the cover plate on the interface
(two M2.5 hex head screws).
3.
Gently lever the connector PCB
(on the end of the cable) out of the socket.
CAUTION: Do not pull the cable to remove
the connector.
4.
Place the connector in an anti-static bag.
5.
Refit the cover plate.