LED diagnostics

Readhead
The readhead set-up LED consists of a tri-coloured LED which can display any combination of Red, Blue or Green.
Signal
Green
Incremental (AGC off)
Orange *
Red
Single-Blue flashing
CAL
Double-Blue flashing
Green (flash)
Reference mark
Blank (flash)
Red (flash)
Restore factory defaults
Four Blue flashes on switch on
When stationary will be Green or Red.
  Flash will effectively be invisible when incremental signal level is > 70% when passing reference mark.
 
When AGC is enabled the LED indication will be as shown but with the addition of a Blue indication.
Ti interface
Signal
Purple
Blue
Green
Orange
Incremental
Red
Red / blank - flashing
Blue / blank - flashing
Purple / blank - flashing
Reference mark
Blank flash
Alarm output will take the form of 3-state or line driven E− signal depending on interface configuration.
Also, some configurations do not output overspeed alarm. See ATOM
-Momentary status only, while fault condition remains.
-Alarm may result in axis position error, re-datum to continue.
NOTE: See
'Troubleshooting', page 31
for more information on diagnosing faults.
ATOM linear installation guide
Indication
Normal set-up; signal level > 70%, AGC off
Acceptable set-up; signal level 50% to 70%, AGC off
Poor set-up; signal may be too low for reliable operation; signal level < 50%, AGC off
Calibrating incremental signals
Calibrating reference mark
Normal phasing
Acceptable phasing
Poor phasing; clean scale and recalibrate if required
Factory defaults restored
Indication
Normal set-up; signal level 110% to 135%
Optimum set-up; signal level 90% to 110%
Normal set-up: signal level 70% to 90%
Acceptable set-up; signal level 50% to 70%
Poor set-up; signal may be too low for reliable operation; signal level < 50%
Poor set-up; signal level < 20%; system in error
Overspeed; system in error
Over signal; system in error
Reference mark detected (speed < 100 mm / s only)
miniature encoder system data sheet (Renishaw part no. L-9517-9563) for more details.
Status
Status
Alarm output
No
No
No
No
No
Yes
Yes
Yes
No
30