Validated Reprocessing Of The Neonatal Flow Sensor - Dräger Evita V Series Reprocessing Instructions

Hide thumbs Also See for Evita V Series:
Table of Contents

Advertisement

3.6.5

Validated reprocessing of the neonatal flow sensor

Preparation
● Remove the sensor insert from the sensor housing before cleaning and
disinfection.
Note the following when reprocessing the sensor insert:
– Do not use any brushes.
– Do not spray the sensor insert.
– Do not clean the sensor insert by machine, and do not disinfect it thermally.
3.6.5.1
Manual cleaning followed by disinfection by immersion
Components:
– Sensor housing of the ISO 15 flow sensor
– Sensor housing of the flow sensor Y-piece
– Sensor insert
Cleaning agent:
neodisher mediclean
forte
Disinfectant:
CIDEX OPA
Prerequisites:
– The cleaning agent and the disinfectant have been prepared in accordance with
the manufacturer's instructions.
Manual cleaning
1. Place the components in the cleaning agent.
2. Swirl the components back and forth at least 3 times. Make sure the cleaning
agent reaches all surfaces and interior spaces.
Reprocessing instructions
Manufacturer
Dr. Weigert
ASP
|
Evita V series / Babylog VN series
Concentration
Contact time
0.5 %
10 min
100 %
5 min
Reprocessing
25