Dräger Evita V Series Reprocessing Instructions page 13

Hide thumbs Also See for Evita V Series:
Table of Contents

Advertisement

Disassembling the expiratory valve or the neonatal expiratory valve
No. Expiratory valve
1
2
3
4
Procedure:
1. Remove the flow sensor sleeve (1) from the expiratory valve or remove the
muffler (3) from the neonatal expiratory valve.
2. Remove the diaphragm (2) and do not disassemble it further.
3. Remove the water trap container (4).
4. Drain the water trap container.
Removing the inspiratory valve from the ventilator
This section addresses how to remove the inspiratory valve from the ventilator
Evita. The neonatal inspiratory valve is removed from the ventilator Babylog in the
same manner.
Prerequisites:
– The inspiratory valve is removed and disassembled only if patient gas has
flowed through the inspiratory valve.
– The device is switched off.
Reprocessing instructions
1
4
Flow sensor sleeve
Diaphragm
-
Water trap container
|
Evita V series / Babylog VN series
2
3
Neonatal expiratory valve
-
Diaphragm
Muffler
Water trap container
Reprocessing
2
4
13

Advertisement

Table of Contents
loading

This manual is also suitable for:

Babylog vn series

Table of Contents