Reprocessing
Disassembling the neonatal flow sensor
Prerequisites:
– The sensor plug was removed from the rear of the device.
Procedure for the ISO 15 neonatal flow sensor:
1. Remove the flow sensor housing (4) from the Y-piece.
2. Disconnect the plug of the flow sensor cable (1) from the flow sensor.
3. Gently press the knobs (2) on both sides while removing the insert (3) from the
sensor housing (4).
Procedure for the neonatal flow sensor Y-piece:
1. Remove the breathing hoses from the flow sensor Y-piece (4).
2. Disconnect the plug of the flow sensor cable (1) from the flow sensor.
3. Gently press the knobs (2) on both sides while removing the sensor insert (3)
from the flow sensor Y-piece (4).
Removing the CO
Prerequisites:
– The sensor plug was removed from the rear of the device.
Procedure:
1. Remove the CO
2. Remove the cuvette (2) from the patient port of the Y-piece.
10
1
1
2
3
4
sensor
2
sensor (1) from the cuvette.
2
1
2
Reprocessing instructions
3
2
4
|
Evita V series / Babylog VN series