15.Check the components for visible damage, paying particular attention to the
measuring wires and their pins, and replace if necessary.
3.6.5.2
Machine cleaning with thermal disinfection
Use a washer-disinfector that meets the requirements of the standard ISO 15883.
Dräger recommends the use of a load carrier for anesthesia accessories and
ventilation accessories. Follow the manufacturer's instructions for the washer-
disinfector.
Components:
NOTICE
► Do not clean the flow sensor insert by machine, and do not disinfect it thermally.
– Flow sensor housing (without flow sensor insert)
Step
Agent
Preliminary clean-
Tap water
ing
Cleaning
neodisher medi-
clean forte
Neutralizing
neodisher Z
Flushing
Demineralized
water
Disinfecting
–
Drying
–
Prerequisites:
– Before machine cleaning with thermal disinfection, manual cleaning (see
"Manual cleaning followed by disinfection by immersion", page 25) must be
performed.
– The washer-disinfector has been prepared in accordance with the
manufacturer's instructions.
Positioning the components in the load carrier
Procedure:
1. Position the components to be stable.
2. Ensure the following:
– All surfaces and interior spaces can be completely rinsed.
– The water can drain off freely.
Reprocessing instructions
Manufac-
turer
–
Dr. Weigert
Dr. Weigert
–
–
–
|
Evita V series / Babylog VN series
Concentra-
Temperature
tion
–
Tap water tem-
perature
Min. 0.2 %
Min. 55 °C
(131 °F)
Min. 0.1 %
Tap water tem-
perature
–
Tap water tem-
perature
–
Min. 90 °C
(194 °F)
–
–
Reprocessing
Contact time
Min. 2 min
Min. 5 min
Min. 1 min
Min. 1 min
Min. 5 min
Drying time
depends on the
load
27