Reprocessing
3.6.7.2
Reprocessing procedures
Manual cleaning followed by disinfection by immersion
Use disinfectants that are nationally approved and are suitable for the particular
reprocessing procedure.
Follow the manufacturer's instructions for the disinfectants.
Components:
– Flow sensor ISO 15
– Flow sensor (Y-piece)
The following disinfectants were compatible with the material at the time of testing:
Disinfectants
Sekusept Pulver CLASSIC
Korsolex Extra, Korsolex Basic
Gigasept FF
Other disinfectants are used at one's own risk.
Components:
– Reusable cuvette for the CO
The following disinfectants were compatible with the material at the time of testing:
Agent
Cleaning agent:
neodisher mediclean forte
Disinfectant:
CIDEX OPA
Other disinfectants are used at one's own risk.
Steam sterilization
The following components can be steam sterilized:
– Reusable cuvette for the CO
– Neonatal flow sensor ISO 15 (sensor insert inserted into the housing)
– Neonatal flow sensor Y-piece (sensor insert inserted into the housing)
The sensor insert must be inserted into the housing during steam sterilization.
Procedure:
1. Sterilize the components (maximum 134 °C (273.2 °F), 5 min).
2. Neonatal flow sensor: Check the components for visible damage, paying
particular attention to the measuring wires and their pins, and replace if
necessary.
3. Reusable cuvette for the CO
and replace if necessary.
30
sensor
2
sensor
2
sensor: Check the components for visible damage
2
Reprocessing instructions
Manufacturer
Ecolab
BODE Chemie
Schülke & Mayr
Manufacturer
Dr. Weigert
ASP
|
Evita V series / Babylog VN series