Flow sensor and expiratory valve
Hide thumbs Also See for Evita V Series:

Advertisement

Quick Links

Dräger Evita
V-Series
®
Flow Sensor and Expiratory Valve
Disassembly/Assembly
Disassembly
/Assembly
Open the flap by lifting the lower
edge upwards.
Close the flap.
Disassembly
/Assembly
Turn the locking ring as far as
possible to the left.
Turn the locking ring as far
as possible to the right
until it clicks into place.
Remove the diaphragm.
Attach the diaphragm onto the
edge of the expiratory valve housing.
Make sure that it fitts properly.
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.
Flow Sensor
Push the flow sensor as far as it
possible go to the left.
Push the flow sensor to
the right into the socket.
Expiratory Valve*
Remove the expiratory valve
from the fitting.
Turn the locking ring as far as
possible to the left, push the
expiratory valve into the fitting.
Remove and empty the water
trap container.
Connect the water trap container.
Remove the flow sensor
from the socket.
Fit the flow sensor
into the socket.
Remove the sensor sleeve
from the expiratory valve.
Attach the flow sensor sleeve
on the expiratory valve.
* The Neonatal Expiratory Valve is
also working with Evita
V-Series.
®
For manual reprocessing of the
Neonatal Expiratory Valve please
refer to Babylog guideline.

Advertisement

loading

Summary of Contents for Dräger Evita V Series

  • Page 1 Dräger Evita V-Series ® Flow Sensor and Expiratory Valve Disassembly/Assembly Disassembly /Assembly Flow Sensor Open the flap by lifting the lower Push the flow sensor as far as it Remove the flow sensor edge upwards. possible go to the left. from the socket.
  • Page 2 Dräger Evita V-Series ® Flow Sensor Reprocessing instructions Manual cleaning Swirl the flow sensor back and forth at least 3 times. Prepare the cleaning and the disinfectant solution in Make sure the cleaning agent reaches all surfaces accordance with manufacturer’s instructions. and interior spaces.
  • Page 3 Dräger Evita V-Series ® Disinfection Machine cleaning Disinfection by immersion Place the flow sensor in the disinfectant. Observe the specified contact time. At the beginning of the contact time, swirl the flow sensor back and forth at least 3 times. Make sure the disinfectant reaches all surfaces and interiors. At the end of the contact time, swirl the flow sensor back and forth again at least 3 times.