Dräger Evita V Series Reprocessing Instructions page 26

Hide thumbs Also See for Evita V Series:
Table of Contents

Advertisement

Reprocessing
3. Fill a syringe that can hold at least 50 mL with the cleaning agent. Remove the
flow sensor housing from the cleaning agent and spray at least 50 mL through
each opening. Put the flow sensor housing back into the cleaning agent.
Sensor housing of the flow sensor Y-piece:
Sensor housing of the ISO 15 flow sensor:
4. At the end of the specified contact time, swirl the components back and forth
again at least 3 times.
5. Fill a syringe that can hold at least 50 mL with the cleaning agent again. Remove
the flow sensor housing from the cleaning agent again and spray 50 mL through
each opening.
6. After the contact time, rinse the components in the water bath (at least drinking
water quality) until no cleaning agent residue is visible.
7. Shake out residual water carefully. Allow the components to dry completely.
8. Check the components for visible soiling and repeat steps 1 through 7, if
necessary.
9. Check the components for visible damage, paying particular attention to the
measuring wires and their pins, and replace if necessary.
Disinfection by immersion
10.Place the components in the disinfectant. Observe the specified contact time.
11. At the beginning of the contact time, swirl the components back and forth at
least 3 times. Make sure the disinfectant reaches all surfaces and interiors.
12.At the end of the contact time, swirl the components back and forth again at
least 3 times.
13.After the contact time, rinse the components in the water bath (at least drinking
water quality) until no disinfectant residue is visible.
14.Shake out residual water carefully. Allow the components to dry completely.
26
Reprocessing instructions
|
Evita V series / Babylog VN series