Reprocessing
5. Disassemble the medication nebulizer in accordance with the corresponding
instructions for use.
3.3.2

Device-specific components

The device-specific components must be removed from the device and, if
necessary, disassembled.
Removing the expiratory valve or the neonatal expiratory valve from the
ventilator
This section addresses how to remove the expiratory valve from the ventilator Evita.
The neonatal expiratory valve is removed from the Evita and Babylog ventilators in
the same manner.
Prerequisites:
– The flap on the front of the ventilator is open.
– The expiratory flow sensor is removed from the ventilator Evita.
Procedure:
1. Turn the locking ring (1) as far as possible to the left.
2. Remove the expiratory valve from the fitting.
12
1
1
Reprocessing instructions
|
Evita V series / Babylog VN series