Figure 4–42. Shield plate in function
6. Close the sliding door. The see-through lid (if in place) protects the
instrument against environmental contamination. However, if
processing potentially biohazardous materials, the see-through lid
must be in place, and the sliding door must be closed.
Note The sliding door can be left open or the see-through lid can be
absent if desired. This action does not break the run.
7. After the run, remove the plate(s) according to the protocol request.
Confirm each plate removal by pressing the
that the plate containing your samples is removed first.
8. Press the
button after completing the run.
STOP
For startup, refer to "Switching on" on page 39.
Figure 4–43. Plates
The PLEX-ID SP is compatible with 24 Deep Well plates and 96 Deep
Well plates (Figure 4–43 and Table 4–4). Specially designed magnetic
rods (magnet heads) and tip combs that protect the magnets during the
process are available for different plate types and applications.
Routine Operation
Shield plate
button. Note
START
PLEX-ID SP User Manual 10/2009 N09455
How to start