Installation
Installation setups
26
PLEX-ID SP User Manual 10/2009 N09455
To remove both the transport locks, follow these steps:
Heating block
transport lock
Figure 3–9. PLEX-ID SP transport locks fitted
1. To remove the transport lock of the tip comb holder, take firmly
hold of the lock and press tightly on both sides of the lock
according to Figure 3–10 so that the lock ends are released. Keep
hold of the tip comb holder and lock simultaneously, and start
gradually lifting the holder.
Figure 3–10. Removing the transport lock of the tip comb holder
2. Lift the tip comb holder higher and continue pressing the transport
lock of the tip comb holder. Remove the transport lock from its
location by ejecting it from the two eccentric pins (Figure 3–11).
When relocating the instrument, refer to "How to refit the
transport lock of the tip comb holder" on page 37.
Tip comb holder
transport lock
Abbott Laboratories