Functional Description

PLEX-ID SP magnetic particle processor

Back / side view

PLEX-ID SP
magnetic
processor
18
PLEX-ID SP User Manual 10/2009 N09455
The back view of the PLEX-ID SP instrument is shown in
Figure 2–4 and the side view in Figure 2–5.
Cooling-air inlet
Figure 2–4. PLEX-ID SP back view
Figure 2–5. PLEX-ID SP side view
The PLEX-ID SP (Figure 2–6) has room for eight plates. The tip
combs are compatible with the plates. During the individual steps, the
plates are kept stationary, and the only moving assembly is the
particle
processing head with tip comb and magnetic rods. The head consists of
two vertically moving platforms. One is needed for the magnetic rods
(24 or 96 pieces) and the other one for the plastic tip comb.
Up to eight plates can be simultaneously on the turntable. However,
during one sample processing, the protocol enables the use of more
than eight plates in total. One tip comb contains 24 or 96 tips used for
processing 24 or 96 samples at a time.
Before starting the magnetic particle processing via the keypad and
display, the samples and reagents are dispensed into the plates and the
tip comb is placed onto a plate, from which it is automatically loaded.
The plates are placed onto the turntable into the corresponding plate
stations according to the protocol instructions. During the operation,
Type label
USB port
Warning marking
Serial port RS-232C
Mains power
supply connector
Cooling-air outlet
ON/OFF switch
Abbott Laboratories