Checking Delivery; Requirements - Abbott PLEX-ID SP User Manual

Table of Contents

Advertisement

Installation

Requirements

Checking delivery

Requirements
24
PLEX-ID SP User Manual 10/2009 N09455
Check the enclosed packing list against order. In case of any deviations,
contact your local Abbott Molecular representative.
Visually inspect the transport package, the instrument and the
accessories for any possible transport damage.
If the carton has been damaged in transit, it is particularly important
that you retain it for inspection by the carrier in case there has also been
damage to the instrument.
Neither the manufacturer nor its agents can be held responsible for any
damage incurred in transit, but the manufacturer will make every effort
to help obtain restitution from the carrier. Upon receipt of the carrier's
inspection report, arrangements will be made for repair or replacement.
If any parts are damaged, contact your local Abbott Molecular
representative.
When you set up your PLEX-ID SP, avoid sites of operation with
excess dust, vibrations, strong magnetic fields, direct sunlight or UV
light, draft, excessive moisture or large temperature fluctuations.
Make sure the working area is flat, dry, clean and vibration-
proof and leave additional room for accessories, cables, and
reagent bottles.
Make sure the ambient air is clean and free of corrosive vapors,
smoke and dust.
Make sure the ambient temperature range is between +5°C
(41°F) and +40°C (104°F).
Make sure relative humidity is between 10% and 80% (non-
condensing).
Leave sufficient space (at least 10 cm, 3.9 in.) on both sides and at the
back of the unit to allow adequate air circulation.
The PLEX-ID SP does not produce operating noise at a level that
would be harmful. No sound level measurements are required after
installation.
Caution Do not operate the instrument in an environment where
potentially damaging liquids or gases are present.
Place the instrument on a normal laboratory table. The net weight of
the entire equipment is approx. 28 kg [62 lbs.].
The instrument operates at voltages of 100 – 240 Vac and the
frequency range of 50/60 Hz.
Abbott Laboratories