Magnetic head
Abbott Laboratories
Table 3–2. Magnet head vs. tip comb
Description
Head for the 24 Deep Well plate
Head for the 96 Deep Well plate
When not in use, keep the magnet heads always in their respective
plastic storage boxes (Figure 3–21).
Figure 3–21. The magnet heads in their storage boxes
Caution The PLEX-ID SP should not be kept in close proximity to
magnetic tapes, computer discs or other magnetic storage systems, such
as credit cards, as these can be damaged by the strong magnetic field of
the magnet heads.
Do not hold the magnet heads close to a PC display, since this may
cause damage to the display.
Do not use metal tools when handling the magnet heads.
Warning This product contains very strong permanent magnets.
People wearing a pacemaker or metallic prostheses should not use this
product. A pacemaker or prostheses may be affected or damaged if it
comes in close contact with a strong magnetic field.
To insert the magnet heads, choose the Change_magnet protocol under
the Maintenance menu by using the Up and Down arrow keys. Then
press
START
black arrows) slot into place (Figure 3–22 and Figure 3–23). When you
remove the magnet head, be careful not to damage the magnetic rods
against the tip comb holder frame.
Tip comb
. Insert the magnet head so that the three pins (shown with
Installation setups
Description
24 Deep Well Tip Comb
96 Deep Well Tip Comb
PLEX-ID SP User Manual 10/2009 N09455