Samples should not be frozen at -20°C.
Samples stored at -25°C for 1 month showed a – 14% bias.
Samples stored at -80°C for 3 months showed a 2% bias.
Do not freeze reagents.
Protect reagents from light.
Do not refreeze calibrator.
The calibrator should not be frozen at -20°C.
The calibrator concentration changes with each lot
number and is listed on the label.
The calibrator is reconstituted with 1 mL DI water,
allowed to stand for 5 minutes, and swirled until
dissolved. Due to the low volume of the calibrator (1 mL),
a 1 mL Class A volumetric pipette should be used.
Do not shake the calibrator. Shaking can cause calibration
inconsistencies due to bubbles in the vial.
The calibrator is traceable to the National Reference
System for Cholesterol, NRS/CHOL.
The calibrator is supplied with the reagent, and cannot be
ordered separately.
®