7 Resend exposures

7.2 Resend images in Rescue mode (recommended)

70
Operating Instructions Xios AE Sensor USB interface and sensors
Resend images in Rescue mode
7.2
(recommended)
To rescue unsaved images, please perform the following steps:
ü Resolve the issue that caused the Sensor plugin to enter Rescue
mode.
1. Click on Windows Start menu > All Programs > Sidexis 4 Sensor
Plugin.
2. Click on Sidexis 4 Sensor Configuration.
3. Click "Devices" at the Sidexis 4 Sensor Configuration page.
4. Select the device in Rescue mode. If several devices are
connected, choose the one in Rescue Mode.
5. The Exposure Information area provides the number of unsaved
images and patient-related information. Click "Resend Images" .
6. When all the images are resent, click "Finish" .
Ä The "Devices" page will now indicate that the system is ready. No
additional recovery steps are needed.
Ä In Sidexis, on the device selection screen, the Rescue icon is
removed. If, for some reason, the images cannot be resent, they
can still be rescued by exporting them from this page and then
importing them manually into Sidexis and saving them in that way.
This alternate procedure is described in the following section.