1 General information
1.6 Demo mode
V
W
X500
12
1.6
Demo mode
Activation
1. Set service routine "26" to "On".
2. Switch the unit off.
3. Open the protective cover of the wall module.
4. Attach cables V (blue) and W (pink) on terminal strip X500.
5. Close the protective cover of the wall module.
6. Switch the unit back on.
The message "E1 11 88" appears on the display.
7. Acknowledge this message by pressing any key (not a release
button).
The demo mode is now activated.
When a release button is activated, no X-ray radiation will be
generated.
Deactivation
✔
The unit is in "demo mode".
1. Set service routine "26" to "Off".
2. Switch the unit off.
3. Open the protective cover of the wall module.
4. Attach cable V (blue) to terminal strip X500.1.
5. Attach cable W (pink) on terminal strip X500.2.
6. Close the protective cover of the wall module.
7. Switch the unit back on.
The demo mode was deactivated.
When a release button is activated, X-ray radiation will be generated.
Dentsply Sirona
Service Manual Heliodent Plus
62 15 102 D3507
D3507.076.01.15.02 03.2018