Posterior Tooth Exposures - Dentsply Sirona Xios AE Operating Instructions Manual

Sensor usb interface and sensors
Table of Contents

Advertisement

4 Operation
4.8 Position sensor with AimRight reusable sensor holder system
52
Operating Instructions Xios AE Sensor USB interface and sensors
4.8.2

Posterior tooth exposures

Preparing the sensor holder
For posterior tooth exposures use the yellow sensor holders.
1. For the right upper jaw and left lower jaw: Fasten the guide rod for
the sensor holder (C) in the perforation (B) of the localizer ring (A).
2. For the left upper jaw and right lower jaw: Fasten the guide rod for
the sensor holder (E) in the perforation (D) in the localizer ring (A).
3. Slide the sensor into the hygienic protective sleeve, see section
"Slide the hygienic protective sleeve over the sensor".
4. Place the sensor on the palm of your hand and clip the sensor
holder onto the sensor.
5. Push the sensor into the sensor holder up to the stop.