Bite wing exposures

For bite wing exposures use the red sensor holder.
1. Place the red localizer ring (B) onto the straight guide rod (C).
2. Place the red sensor holder tab (A) onto the guide rod (C).
3. Slide the sensor into the hygienic protective sleeve, see section
"Slide the hygienic protective sleeve over the sensor".
4. For vertical bite wing exposures: Glue the sensor holder tab onto
the sensor's hygienic protective sleeve. Align the tab vertically to
the sensor and place it in the center on the active sensor surface,
as shown in the diagram.
5. For horizontal bite wing exposures the tab must be placed aligned
horizontally to the sensor. See the adjacent drawing.
6. Position the sensor in the patient's mouth.
7. Bring the X-ray tube assembly into the correct position and take an
X-ray exposure.
8. Remove the sensor from the hygienic protective sleeve. For this
follow the instructions in section "Remove the hygienic protective
sleeve from the sensor". The used sensor holder tab and hygienic
protective sleeve must be disposed of after the examination.
9. Clean and sterilize the guide rod and localizer ring.