7.3.15. Display Q2 sensor values

The list of sensor values can be opened by clicking on this symbol.
Generally, this display is only used for service purposes and for locating errors.
Figure 7.41:
Q2 sensor values display
A complete list of all sensor values is displayed in the following table:
Sensor signal
Use/Meaning
UIR 1
For internal calculations only
UIR 2
For internal calculations only
UIR 3
For internal calculations only
UIR 4
For internal calculations only
PI
Pressure (ambient + measuring unit [approx.
160 mbar])
TSB
Temperature of sensor block
UW11
For internal calculations only
UW12
For internal calculations only
UW21
For internal calculations only
UW22
For internal calculations only
PU
Ambient pressure (internal monitoring)
TP
Circuit board pressure (internal monitoring)
PW1
For internal calculations only
PW2
For internal calculations only
TM1
For internal calculations only
TM2
For internal calculations only
TB1
For internal calculations only
TB2
For internal calculations only
TH1
Housing temperature (internal monitoring)
Information for general use
Rev. M / 73023639
GASLAB Q2 DISPLAYS (OPERATE THE DEVICE)
126
Sensor value range
Several 100 mV
Ambient-dependent
approx. 70°C ≙ 158°F ≙ 343K
Several 100 mV
Ambient-dependent
Ambient-dependent
Several mW
Around 100°C
Around 68°C
Ambient-dependent
GasLab Q2