Connect The Calibration Gas - Honeywell GasLab Q2 Manual

Gas quality analyzer
Hide thumbs Also See for GasLab Q2:
Table of Contents

Advertisement

5.3.4. Connect the calibration gas

For proper operation of the GasLab Q2 a regular automatic calibration with a binary mixture of
methane (CH
) and carbon dioxide (CO
4
cylinder close to the instrument.
Prerequisite: A suitable pressure reduction and shut-off devices are required to connect the gas
cylinder to the GasLab Q2 via the permanently installed
If an automatic verification measurement is to be carried out via the calibration gas inlet, an
additional automated 3-way valve must be installed in the gas flow. Only clean, particle-free gas can
be used at this inlet.
Complete the following steps in accordance with national regulations on the basis of the
information in this document.
aux normes locales en respectant les instructions de ce manuel.
Refer to /
Failure to do so may result in significant destruction of the device or the system.
respect de ces étapes peut entraîner des dommages importants sur l'appareil ou
l'installation!
Step
Action: Prepare the calibration gas line connecting the device
Connect the end of the line intended for the device connection to the vent gas
1.
properly using a hose.
Carefully open the shut-off valves to create a permitted gas flow above the operating
2.
conditions. Then flush for around 30 seconds to clean the pipeline system.
Close the line shut-off again. Set to a process gas pressure between
0.15 MPag ≙ 21.76psig ≙ 1.5barg and maximum 0.25 MPag ≙ 36.26psig ≙ 2.5barg.
3.
(The exact pressure will be set later).
The line is now ready for connection to GasLab Q2 which should be done immediately.
Therefore, remove the protection (plugs / screw plugs) from the (CAL) inlet of the
4.
instrument and connect the gas line - according to figure 5.2.
Carefully open all the shut-off valves in the gas path so that the gas is present at the
5.
device inlet
Check the tightness of the calibration gas path from the source (gas cylinder) to the
6.
device
10.1.10 System tightness test
Information for general use
Rev. M / 73023639
GASLAB Q2 AFTER DELIVERY AND AT THE PLACE OF USE
) is necessary. Normally the mixture is taken from a gas
2
11.3 List of suitable gases and calibration gases
Honeywell can supply prefabricated boards
with high pressure reduction systems for this purpose.
Suivez impérativement les étapes suivantes conformément
Respectez les
10.3 Connecting and replacing gas cylinders
lines.
63
Le non-
GasLab Q2

Advertisement

Table of Contents
loading

Table of Contents