Honeywell GasLab Q2 Manual page 183

Gas quality analyzer
Hide thumbs Also See for GasLab Q2:
Table of Contents

Advertisement

Refer to the information in the online help which contains more details required to
Ethernet settings
For data communication, connections via Ethernet are mainly used. Go to parameter branch:
GasLab Q2 – Basic system – I/O – Board 0: CPU. Fill in your changes, proceed like explained in the
online help.
Serial interface RS485
These connections include, for example, other measuring instruments, devices for final processing
and evaluation, PLC systems, etc. Go to parameter branch GasLab Q2 – Basic system – I/O – Board
0: CPU. On the "Parameters" tab: Set the type of CH1
TB4-4 to TB4-6)
from "Unused" to "Protocol channel".
Digital inputs
There are two electrically isolated inputs supplied by the device. Go to parameter branch GasLab
Q2 - Basic system – I/O – Input channels and set the type from "Unused" to "Message input" or
"LF pulse input".
Digital outputs
There are two outputs (electrically isolated passive output circuits). Go to parameter branch
proChain GC – Basic system – I/O – Output channels set the type from "Unused" to to the desired
function offered in the pop-up menu e.g.
Analog outputs
Four common galvanically isolated active output circuits ("common ground" / short-circuit proof)
with 0 or 4 to. 20 mA are available. Go to the parameter branch
output channels
"and set in the register card the type of current output for the relevant channel (in
the range I1+ I-, I2+ I-, I3+ I- or I4+ I-). For the Current Range parameter, set the desired current
output range: - 0 to 20 mA or 4 to 20
Information for general use
Rev. M / 73023639
CONFIGURATION AND ANALYSIS SOFTWARE ENSUITE
define all the parameters for the following settings.
"Message output" or "Pulse
mA.
183
(terminals TB4-1 to TB4_3)
output".
"GasLab Q2 – basic system I/O -
or CH2
(terminals
GasLab Q2

Advertisement

Table of Contents
loading

Table of Contents