8.5.

Remote operation panel functions of enSuite

The so-called remote operation panel refers to a function that enables the remote control of enCore
devices with your service computer. The remote operation panel is a copy of the local operation
panel and is displayed in enSuite. The remote access can be done via Ethernet or USB connection.
Since the basic system 03-39-A, the connection is established via MMS and secured via TLS.

8.5.1. Open and close remote operation panel

As soon as you are connected to a device (see
the entries in the lower left corner of the enSuite interface are activated. If you select the "Remote
operation panel" entry there, a graphic representation of the front of the device will appear in the
middle of the enSuite interface. By mouse clicks in the areas of the 7 touch fields (see figure 7.1),
the same operation as on the device is possible. The screen and LED displays are now synchronized
with the device. At the same time, the physical local display is assigned a lock screen and local input
is deactivated.
Figure 8.43: View of the remote operation panel via enSuite
see
8.2.2 Starting enSuite; the first steps
After deactivating the function by closing the window with the X in the first line (shown in the figure
above), by disconnecting the device (see section
Information for general use
Rev. M / 73023639
CONFIGURATION AND ANALYSIS SOFTWARE ENSUITE
8.2.4 Establish a connection to the enCore
Tip: Detach the window from the enSuite frame
displayed enlarged, which makes it easier to use
8.2.5 Disconnecting the device and Connection
207
and have the remote operation panel
device),
.