Ethernet
Cable type refer to 
5.4.2 Cables (power supply / communication)
The connection is made using the terminals designated TB3-1 to TB3-4 in the Q2 connection box.
The core colors depend on the standard used, normally EIA/TIA 568B. The twisted pairs must not be
untwisted or regrouped. The polarity of the cores within the core pair is unimportant and the core
pairs can even be swapped.
Color depending on standard
Core
Plug
EIA/TIA 568 B
pair
(pins)
1
4 / 5
blue + white/blue
2
3 / 6
white/green
+ green
3
1 / 2
white/orange
+ orange
4
7 / 8
white/brown +
brown
The following drawing is only designed to illustrate the screening using an example.
Figure 5.7: Sample LAN/Ethernet connection (diagram)
Additional to the wiring, settings must be made before this interface can be used.
On the Q2
Or with enSuite 
Information for general use
Rev. M / 73023639
GASLAB Q2 AFTER DELIVERY AND AT THE PLACE OF USE
EIA/TIA
IEC
568 A
blue +
white/+
white
blue
white +
red+
orange
orange
white +
black+
green
grey
white +
yellow+
brown
brown
System display I/O (basic network settings / Inputs /Outputs)
7.3.14
Changing the IP address basic network and I/O
8.4.5
length 100m ≙ 328ft.
REA
DIN 47.100
white+
white+
blue
brown
turquoise
green+
violet
yellow
white+
grey +
orange
pink
turquoise+
blue +
violet
red
74
Terminal
Remark
inside
Q2
TB3-3
TB3-4
Only core
pairs 2 and 3
TB3-1
are required
TB3-2
settings".