Logging Of Device Events With Logbooks - Honeywell GasLab Q2 Manual

Gas quality analyzer
Hide thumbs Also See for GasLab Q2:
Table of Contents

Advertisement

2.8.5. Logging of device events with logbooks

Log files can provide useful information to help troubleshooting on site, but also provide
valuable data on the use of the device and any violations. Please note therefore the
contents of these files.
aider au dépannage sur place, mais ils peuvent également fournir des données précieuses
sur l'utilisation de l'équipement et sur les éventuelles violations. Veuillez donc noter que le
contenu de ces dossiers
In enCore devices, a distinction is made between the general audit trail and the fiscal audit trail.
Both are log archives:
General audit trail: The audit trail is a fixed component of the Basic System and exists in every
enCore device. All parameter changes and the following system events are logged here:
The audit trail has a depth of 1000 entries. As soon as the logbook is full and another event occurs,
the oldest entry is overwritten by the latest entry using the FIFO (first-in-first-out) principle.
Fiscal audit trail: enCore devices which are used for legal metrology purposes also have a fiscal
audit trail. All changes of parameters made while the security switch is closed, which are labeled
with the
symbol in the device parameterization, are logged here.
The fiscal audit trail has a maximum storage capacity of 1000 entries. As soon as the fiscal audit
trail is full, parameters with the relevant symbol can no longer be changed while the security switch
is closed – the fiscal audit trail must first be cleared. This action can only be performed by an
authenticated user with the opened security switch.
Information for general use
Rev. M / 73023639
SAFETY AND WARNING INFORMATION
Les fichiers journaux peuvent fournir des informations utiles pour
System messages
All parameter changes
User login and logout
Setting counters
All entries in the fiscal audit trail
The sections on operation and parameterization
contain further information on handling the logbooks
40
GasLab Q2

Advertisement

Table of Contents
loading

Table of Contents