Error Messages - ABB Platinum standard Series Product Instruction

Smart level transmitter
Table of Contents

Advertisement

TROUBLESHOOTING
Message
Calibration required
Cannot save, configura-
tion too large
Cell characterization
board EEPROM failure
Cell temperature over
alarm
Cell temperature under
alarm
Cell temperature over
limit
Cell temperature under
limit
Config "tagname" already
exists
Damaged cell or charac-
terization board
Dynamic temperature
measurement failure
Electronics temperature
measurement failure
Electronic temperature
out of range
Error! Attempt to config-
ure duplicate address
ERROR MESSAGES AND CORRECTIVE ACTION
6 - 2
Table 6-1. Error Messages
Fail
Probable Cause
Mode
No
Cell not calibrated with
present amplifier board.
No
Transmitter type not sup-
ported by Type STT terminal
firmware revision level.
Yes
Damage to cell characteriza-
tion board.
No
Cell temperature under/over
alarm limit.
No
Cell temperature input
exceeds upper factory limit.
No
Cell temperature input
exceeds lower factory limit.
No
Attempted to copy configura-
tion to cartridge or transmitter
where it already exists.
Yes
Cell or characterization board
disconnected.
Cell or characterization board
damaged.
No
Dynamic temperature input
exceeded limits set at factory.
No
Hardware failure of on-board
temperature sensor.
No
Temperature inside electron-
ics housing exceeds limit set
at factory.
No
Attempt was made to assign
analog mode transmitter
address on FBM.
Attempt was made to assign
digital mode transmitter to
occupied FBM address.
Corrective Action
Calibrate transmitter.
Consult Bailey Controls Company
regarding latest Type STT terminal
firmware revision.
Turn Type STT terminal off and then
on. If error still exists after reset,
replace cell and characterization board
assembly
(Section
8).
Correct temperature problem.
Change alarm value. Refer to Type
STT Smart Transmitter Terminal
Instruction .
Remove source of temperature
extreme.
Remove source of temperature
extreme.
Erase old configuration before copying
new one. Refer to Type STT Smart
Transmitter Terminal Instruction .
Make sure cell, characterization board,
and amplifier assembly are mounted
properly.
Replace cell and characterization
board assembly
(Section
8).
Output may no longer be accurate.
Remove source of temperature shift.
Replace amplifier assembly
(Section
8).
Remove source of temperature
extreme.
Be sure transmitter is in correct mode
(analog or digital). Use VIEW key on
Type STT terminal to verify mode.
Refer to Type STT Smart Transmitter
Terminal Instruction for details.
Assign transmitter to unoccupied
address or channel number of FBM.
WBPEEUI110504A0