Agency
FM
Explosionproof
Dust-ignitionproof
Intrinsic safety
Nonincendive
CSA
Explosionproof
Dust-ignitionproof
Intrinsic safety
Nonincendive
Type PTSDD Differential Pressure Transmitter Specifications
Table 1-5. Type PTSDD Differential Pressure Transmitter Specifications
Property
Measurement range,
turndown ratio, zero
suppression, and zero
elevation
WBPEEUI110504A0
Table 1-4. Ambient Condition Limits for Certified Applications
T
max
Category
°C
80
65 (T4)
60
40 (T3C)
104 (T3C)
60 (T3A)
60 (T4A)
SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE.
Table
1-5
ential pressure transmitters.
Lower range value (zero) and upper range value (100%) can be calibrated at any
value of pressure, provided that:
1. Their algebraic difference (the calibrated span) corresponds to an
authorized TDR.
2. Both are within the following acceptable limits.
Range Limits
Type
(kPa)
Upper Lower
PTSDDA
1.5
PTSDDB
7.5
PTSDDC
30
PTSDDD
90
PTSDDF
500
PTSDDG
1,800
PTSDDH
6,000
T
min
°F
°C
°F
kPa
176
-40
-40
108
149 (T4)
140
140 (T3A)
140 (T4A)
lists the specifications unique to Type PTSDD differ-
Characteristic/Value
International System Units
Span and Turndown Ratio (TDR)
Nominal (kPa)
Maximum (kPa)
-1.5
1.5
1:1 TDR
3
-7.5
7.5
1:1 TDR
15
-30
30
1:1 TDR
60
-90
90
1:1 TDR
180
-500
500
1:1 TDR
1,000
-1,800 1,800 1:1 TDR
3,600
-6,000 6,000 1:1 TDR
12,000 0.5:1 TDR
INTRODUCTION
P
P
O
max
min
psia
kPa
psia
15.7
86
12.5
Minimum (kPa)
0.5:1 TDR
0.1
15:1 TDR
0.5:1 TDR
0.4
20:1 TDR
0.5:1 TDR
1.5
20:1 TDR
0.5:1 TDR
4.5
20:1 TDR
0.5:1 TDR
25
20:1 TDR
0.5:1 TDR
90
20:1 TDR
400
15:1 TDR
SPECIFICATIONS
1 - 11
2max
%
21