Chapter 3. Creating Data Recorder Files
Reading Type
Number of Elevations Quantity of elevations.
Elevation Label
Length
Points Per Tube
Point Label Length
Boiler File
Starting Tube #
Ending Tube #
File Name
Units
T-MIN
T-MAX
70
Your Complete Source for
Testing Equipment. Since 1969!
Table 3: Items to be Defined for Each File Type (cont.)
Set file style to Standard (normally recommended) or Extended.
Sets the maximum length of the label identifying each elevation.
Defines the number of points to be recorded for each tube.
Sets the maximum length of the label identifying each elevation.
Defines the number assigned to the first tube in the data file.
Defines the number assigned to the last tube in the data file.
16 character file name, set using the virtual keyboard. No two files in the DMS Go memory
may have the same file name.
Selects either inches or millimeters as the unit of measurement
Defines the minimum thickness limit. Measurements below this limit cause an alarm to signal.
Defines the maximum thickness limit. Measurements above this limit cause an alarm to signal.