Title of Test
Receiver
Gain Control
Frequency ranges
Lower frequency ranges
Upper frequency ranges
Center frequency
Dynamic range
Performance
Minimum and Maximum
Measurable Thickness
Minimum and Maximum
Measurable Thickness
Minimum and Maximum
Measurable Thickness
Minimum and Maximum
Measurable Thickness
DMS Go Operator's Manual
Your Complete Source for
Testing Equipment. Since 1969!
Table 10: EN15317 Specifications (cont.)
Dimension
Min
MHz
Broad Band
MHz
0.963
1.07
MHz
18.495
20.55
MHz
4.432
4.665
dB
-
> 105
mm
1.2
mm
5
mm
0.6
mm
0.8
www.BergEng.com
Berg Engineering & Sales Company, Inc.
Typical
Max
Automatic
1.177
22.605
4.898
-
650
650
75
300
Appendix D. EN15317 Specifications
Comment
with DA501
with DA503
with DA512
DA590
157
1-847-577-3980